Compared to the measuring devices already available on the market, the new "Interdendritic Feeding Measurement" provides additional information about the feed capacity of a melt and prevents casting defects such as blowholes, sink marks and hot cracks.
The SA800SN divides the feeding types into 5 phases and calculates the percentage feeding degrees in the transition phase.
The volume contraction in the transition region from the liquid to the solid phase occurs abruptly in most casting alloys and can lead to various casting defects.
These influencing factors are reflected in different types of feeding and can be divided into: Liquid feeding, Mass feeding, Interdendritic feeding, Burst feeding and Solid feeding.
Further information:
With the device SA800SN, both a feed analysis and a "classical" thermal analysis (for the determination of grain refinement and modification, see e.g. TA750SN) can be carried out.
The system records the cooling process of a metal sample in detail and analyses it automatically. The temperature acquisition and measurement results are graphically displayed.
Tabular overview of the current measurements with sample identification and measurement result.
User-related identification of each measurement (e.g. "article", "customer", "batch", "furnace", "alloy", etc.) with plausibility check
Self-configurable selection or validation lists can simplify sample identification and avoid input errors.
Automatic measured value identification for frequently recurring entries, e.g. for follow-up measurements
Automatic recording and analysis of the cooling process of the metal sample with simultaneous graphic display
Detailed graphic display of the measurement result with indication of the points with thermal relevance. This enables verification of the measurement result.
Search and filter function for all stored measurements
Feeding-Analysis measurements can be subsequently supplemented with additional data (e.g. VAC-TEST / spectral / temperatures) and a comment
If a Density Index Analysis System is connected to this device, VAC-Test measurements can also be documented separately (without a feeding-analysis measurement)
Printout of measurements on a local or network printer (if available)
Customised software adaptations are of course possible
Secured system ("kiosk mode"). No access to the operating system for the operator
Manual backup of measurements on USB data carrier
Further individual functionalities are possible when connected to a network, e.g.:
Calibrated temperature reading acquisition with 60 measurements per second
Evaluation range from 720 - 450 °C (changes possible)
Highly accurate temperature resolution of 0.1 K
Automatic sensor system: Automatic raising and lowering for safe and simple temperature recording
Fanless industrial PC: Intel ATOM CPU, SSD, Windows 10 IoT LTSC
Compact keypad with wipeable dust protection membrane
15" LED monitor
Interfaces: USB, Serial D-Sub, Network RJ45
Fan drawer for cooling the sample ingots to room temperature
Can be used as a table-mounted device or a compact mobile device
Side casting ladle rack included
CE conformity
Tested according to DGUV provisions
We provide a 24-month warranty. The prerequisite for this is annual servicing by IDECO GmbH. (12-month warranty with 3-shift operation)